UV/Visible Light Intensity Meters |
|
UVICURE®
PLUS II and UV POWER PUCK® II
(NEW)
Features:
- Measures UV Peak irradiance (microW/cm2,
mW/cm2, W/cm2)
and total energy density (microJ/cm2,
mJ/cm2, J/cm2)
of one spectral bandwidth (Uvicure
Plus II) or 4 different bandwidths
simultaneously (UV Power Puck II).
- New graph, data, and reference
modes to provide choices in how to
view the collected data
- Compact Size: 4.6" Diameter
(117mm) x 0.5"H (13mm)
- Self-contained - battery-powered
(user-replaceable AAA batteries)
- Unit communicates via cable to
your PC
|
Applications:
- Curing applications of inks,
adhesives, solder masks, and epoxies
(not for use in web press
applications)
- Measure lamp performance
- Compares the efficiencies of
different UV curing systems
- Compare any current reading to
chosen baseline reference to compare
your system over time
- Establish UV level for proper
curing
|

Introduction
The radiometers that first set the standard
for the UV industry are now setting a new
standard with advanced features and an easy to
read display, multiple user selectable modes,
and PC communication for data logging and
trending capabilities, and process verification.
The new EIT Instrument Markets UVICURE Plus
II and UV Power Puck II are advanced versions of
the UVICURE Plus and UV Power Puck widely used
throughout the global UV industry. With user
selectable sample rates, these new instruments
can be used for fast conveyor lines or slower
lines, and measurements are compatible with
other EIT products.
Standard Features and Benefits Include:
Easy to Use. One-Button Operation for On/Off
and Run

Easy to Read Data Display of All 4 Bands.
Data collected for all 4 bands on the UV Power
Puck II is displayed simultaneously on screen in
mW/cm2 and mJ/cm2 for quick and easy viewing by
the operator. No need to toggle through all
eight readings, one screen at a time. Soft
buttons are used for function selections, and
are indicated on the bottom of the display for
easy operator selection and use.
Standard EIT Multiple Bandwidths:
UVA (320-390nm), UVB (280-320nm), UVC
(250-260nm), UVV (395-445nm)
Dynamic Range
Standard unit - 10 Watt; UVA, UVB, UVV; 1 Watt
UVC. Low Power unit -100 mW
Setup Function
Provides user selectable instrument default
modes for data analysis and comparison, screen,
and operational settings.

Reference Mode
Used for comparison between readings. Can be
useful for system setup and troubleshooting. The
user can store the selected UV reading in the
radiometer as a base line or reference reading,
then compare that reading to another. The
radiometer will display both readings and
indicate the percentage of change between
readings. Data is displayed in mJ/cm2 and mW/cm2,
and percentage.

Graph Mode
A graph illustrating the collected UV irradiance
and energy is displayed for each of the UV
bands. Data is expressed in mW/cm2 vs. time.
Graphs at right illustrate both a one and
two-lamp system.
User Selectable Sample Rate
Smooth On Data: Compatible with previous Power
Puck versions.
Smooth Off Data: Compatible with UV PowerMAP at
over 2000 samples per second.

Unit of Measure
The unit of measure is user selectable to
provide ease of reading for operators. Display
the data as you want to see it. Selections are:
mJ/cm2, mW/cm2, J/cm2,
W/cm2, µJ/cm2, µW/cm2
Colorful, Easy to Read Display
Select low, medium, or high intensity for the
graphical display.
Communications Port
Serial communication protocol between unit and
PC/PDA. Download collected data to a computer
for statistical analysis and data logging, and
process verification.
SPECIFICATIONS:
|
UVICURE PLUS II and UV POWER PUCK II
|
|
Display |
Easy to Read, Yellow Text on Black
Background |
|
Range |
10
Watt: UVA, UVB, UVV - 10mW/cm² to
10W/cm²; UVC - 5mW/cm² to 1W/cm² Low
Power Versions: UVA, UVB, UVC, UVV:
100microW/cm² to 100mW/cm² |
|
Accuracy |
+/- 10%; +/- 5% typical |
|
Spectral Ranges (UV Power Puck® II) |
4-channel simultaneous and continuous
monitoring. 320-390nm (UVA), 280-320nm (UVB),
250-260nm (UVC), 395-445nm (UVV) |
|
Spectral Ranges (UVICURE® Plus II) |
1-channel continuous monitoring.
320-390nm (UVA), 280-320nm (UVB),
250-260nm (UVC), 395-445nm (UVV) |
|
Spatial Response |
Approximately cosine |
|
Operating Temperature |
0-75°C Internal temperature; tolerates
high external temperatures for short
periods (audible alarm indicates when
temperature has exceeded tolerance) |
|
Time-Out Period |
2
minutes DISPLAY mode (no key activity). |
|
Battery |
Two user-replaceable AAA Alkaline Cells |
|
Battery Life |
Approx. 20 hours with display on |
|
Dimensions |
4.60 x 0.50 inches; 117 mm x 12.7 mm (D
x H) |
|
Weight |
10.1 ounces (289 grams) |
|
Package Material |
Aluminum, stainless steel |
|
Carrying Case Material |
Cut polyurethane interior, scuff
resistant nylon exterior cover |
|
Carrying Case Weight |
9
ounces (260 grams) |
|
Carrying Case Dimensions |
10.75 x 3.5 x 7.75 inches; 274 x 89 x
197 mm (W x H x D). |
*
Specifications are subject to change. EIT will
not be responsible for discrepancies due to
typographic errors.
|
|
|
 |
UVICURE PLUS
- Used in curing applications of inks, adhesives, soldermasks,
and epoxies (not for use in web press applications)
- Incorporates one UV spectral response range
- Display provides readout of total UV dosage in Joules/cm2
and peak UV intensity in Watts/cm2
- Measures UV lamp performance
- Compares the efficiencies of different UV curing systems
SPECIFICATIONS:
| UVICURE
PLUS® RADIOMETER |
| Range |
UVA,
UVB, UVV - 5mW/cm2 to 5W/cm2
UVC - 5mW/cm2 to 1W/cm2 |
| Display |
4
digit LCD |
| Data
Readings |
UV
Bandwidth: UVA (320-390nm), UVB (280-320nm), UVC
(250-260nm), or UVV (395-445nm)
Total Energy 0 to 250 Joules/cm2
Peak Intensity:
UVA, UVB, UVV - 0 to 5W/cm2
UVC - 0 to 1W/cm2
Internal Radiometer Temperature Degrees Celsius (80 C
max.)
Lo Batt Indicator |
| User
Interface |
Push
button switch allows user to obtain data from both display
modes, intensity and dosage |
| Accuracy |
+/-5%
typical, +/-10% guaranteed |
| Spectral
Response |
Choose
320-390nm (UVA), 280-320nm (UVB), 250-260nm (UVC), or
395-445nm (UVV) |
| Spatial
Response |
Approximately
cosine |
| Operating
Temperature Range |
0-80
C internal. The UVICURE Plus will tolerate much higher
external temperatures for short periods. A safety audible
alarm will sound if internal temperature rises above a
safe level. |
| Time-Out
Period |
Approximately
4 minutes RUN mode; 2 minutes DISPLAY mode |
| Batteries |
Two
user replaceable lithium cells, Duracell DL2450, Sanyo CR
2450 or equivalent |
| Battery
Life |
Over
1500 readings with typical use |
| Dimensions |
4.60"
diameter x .50" high (11.68cm x 1.27cm) |
| Weight |
11.75
oz. (333.11 grams) |
| Package
Material |
Stainless
and aluminum |
| CARRYING
CASE |
| Construction |
Cut
polyurethane interior. Soft, scuff-resistant nylon
exterior cover |
| Weight |
1
lb. (453.6 grams) |
| Dimensions |
12"W
x 4.7"H x 8.25"D (30.48 x 11.94 x 20.96cm) |
|
|
 |
UV POWER PUCK
- Used in curing applications of inks, adhesives, soldermasks,
and epoxies (not for use in web press applications)
- Incorporates four separate spectral response ranges and
measures all four simultaneously
- Display scrolls through UV dosage in Joules/cm2 and peak UV
intensity in Watts/cm2 of all four separate UV spectral
response ranges
- Measures UV lamp performance (and possible bandwidth shift
for doped lamps)
- Compares the efficiencies of different UV curing systems
SPECIFICATIONS:
| UV
POWER PUCK® RADIOMETER |
| Range |
UVA,
UVB, UVV - 5mW/cm2 to 5W/cm2
UVC - 5mW/cm2 to 1W/cm2 |
| Display |
4
digit LCD |
| Data
Readings |
UV
Bandwidth:
UVA (320-390nm)
UVB (280-320nm)
UVC (250-260nm)
UVV (395-445nm)
Total Energy 0 to 250 Joules/cm2
Peak Intensity:
Peak Intensity:
UVA, UVB, UVV - 0 to 5W/cm2
UVC - 0 to 1W/cm2
Internal Radiometer Temperature Degrees Celsius (80 C
max.)
Lo Batt Indicator |
| User
Interface |
Push
button switch allows user to obtain data from all display
modes in each of the four spectral ranges |
| Accuracy |
+/-5%
typical, +/-10% guaranteed |
| Spatial
Response |
Approximately
cosine |
| Operating
Temperature Range |
0-80
C internal. The UV POWER PUCK will tolerate much higher
external temperatures for short periods. A safety audible
alarm will sound if internal temperature rises above a
safe level. |
| Time-Out
Period |
Approximately
4 minutes RUN mode; 2 minutes DISPLAY mode |
| Batteries |
Two
user replaceable lithium cells, Duracell DL2450, Sanyo CR
2450 or equivalent |
| Battery
Life |
Over
1500 readings with typical use |
| Dimensions |
4.60"
diameter x .50" high (11.68cm x 1.27cm) |
| Weight |
11.75
oz. (333.11 grams) |
| Package
Material |
Stainless
steel and aluminum |
| CARRYING
CASE |
| Construction |
Cut
polyurethane interior. Soft, scuff-resistant nylon
exterior cover |
| Weight |
1 lb.
(453.6 grams) |
| Dimensions |
12"W
x 4.7"H x 8.25"D (30.48 x 11.94 x 20.96cm) |
|
|
 |
SPOTCURE
- Monitors the output of a UV "spot curing" system
- Provides indication of light guide degradation and spot
curing system degradation
- Measures intensities up to 19.99 Watts/cm2
- Includes several different size light guide adaptors for
maximum flexibility
SPECIFICATIONS:
| SpotCure®
RADIOMETER |
| Intensity
Range |
0-19.99W/cm2 |
| Resolution |
10mW/cm2 |
| Accuracy |
+/-5%
typical, +/-10% maximum |
| Display |
3 1/2
digit LCD |
| Power
Source |
Lithium
Battery Stick |
| Battery
Life |
12,500
hours continuous operation (over 100,000 readings) |
| Overall
Dimensions |
6.40"L
x 1.74" Diameter (16.26cm x 4.42cm);
Measurement Head - 2.13" Diameter (5.41cm) |
| Weight |
12.8
oz. (358 grams) |
| Operating
Temperature |
0-70°C |
| Display
Time |
2.5
to 5 minutes |
| Spectral
Ranges |
Choose
320-390nm (UVA); 280-320nm (UVB); 250-260nm (UVC); or
395-445nm (UVV) |
| Case
Materials |
Aluminum,
polyester, quartz |
|
|